Mechanistic studies of tissue injury in SWL


1. Single bubble oscillation inside silicone tubes
To confirm the tissue constraining effect on bubble expansion in vivo, we have performed an experiment on single bubble oscillation in silicone tubes, using a Nortech electrohydraulic lithotripter (EHL, 1.9F microprobe). The tube inner diameter ranges from 1.5 mm to 3.5 mm. High-speed imaging system and focus hydrophone were used during the experiment.

Fig. 1 shows the representative high-speed sequences of EHL-induced bubble oscillation both in water and inside silicone tubes at an output setting of 70%. Following each spark, a cavitation bubble was produced and expanded freely in water to a maximum size of 5 mm and then collapse. In comparison, the expansion of an EHL-induced bubble inside a 1.5 mm and 3.5 mm silicone tubes were significantly constrained by the tube wall, leading to a shortened expansion of the bubble and more rapid collapse. It is interesting to note that the smaller the tube size, the larger the relative dilation of the tube. However, no rupture of the silicone tubes was observed after 100 shocks. The AE signals measured simultaneously revealed the significant reduction in the collapse time of the bubble inside the silicone tubes compared with that in water (see Fig. 2a). However, in contrast to the strong collapse in water, the secondary peak pressure from bubble collapse inside the silicone tubes was greatly diminished because of the significantly reduced bubble expansion (see Fig. 2b).

2. Shock wave-induced bubble oscillation in tissue phantom and mechanism of vessel rupture

To investigate the vascular injury during SWL, silicone tubes of 500 mm inner diameter and cellulose hollow fibers of 200 mm inner diameter were used as tissue phantom. High speed imaging system (see Fig. 3) was also set up to visualize the shock wave-induced bubble dynamics inside silicone tube and hollow fiber. Fig. 4 shows the bubble oscillation in silicone tube and hollow fiber and the rupture of the fiber. It can be seen that the expansion of the shock wave-induced bubbles was significantly constrained by the tube wall. The smaller the tube the shorter the collapse time of the bubble would be. On the other hand, the expansion of the bubbles inside the tube also cause significant dilation of the tube wall. In our experiments, no damage was observed in the silicone tubes after several hundred shocks. This is partially attributable to the large wall thickness (300 to 400mm) of the silicone tubes and partially to their intrinsic tensile failure strength (sf = 8.4 MPa) which is much higher than that of small blood vessels (sf= 1.47 ~ 5.07 MPa). However, for the hollow fiber (3 mm wall thickness), only about 20 shocks generated by the XL-1 lithotripter at 16 KV were needed to cause a rupture, with circulating fluid inside the hollow fiber leaking out into the surrounding caster oil (Fig. 4c). The number of shocks needed to cause a rupture varied with the output voltage of the lithotripter (Fig. 5). The increased propensity for rupture at a higher KV (thus a potentially larger bubble expansion) indicated that the ratio between the potential bubble expansion and vessel size is an important determinant for the rupture. Interestingly, when the pressure waveform of the XL-1 generated shock wave was inverted by a pressure-release reflector insert to suppress the large intraluminal bubble expansion, no rupture of the hollow fibers could be observed up to 100 shocks at 20KV. These results strongly support our hypothesis that large intraluminal bubble expansion could rupture capillary and small blood vessels if adequate cavitation nuclei exist in blood.

See following two papers for more and detail information.


Back to Dr. Songlin Zhu's Homepage